» Calcium Flow Plus ISE
The directION calcium combination ISE has a solid state PVC polymer membrane with an integral liquid reference system. The electrode is designed for the detection and analysis of Calcium ions in aqueous solutions and is suitable for use in the laboratory and as a component in on-line analysers.
The Flow plus range incorporates two solid state ISE’s in one 12 mm Epoxy body. One acts as the Sensor whilst the other is incorporated into the reference system. Your Flow plus combination ISE therefore does not require a separate reference electrode, making it convenient to use with small sample volumes. The reference half cell has a free-flowing liquid junction which will reduce drift and provide more stable readings than conventional ISE’s. Flow plus electrodes are ideal for measuring awkward or dirty samples, which would clog conventional reference electrodes.
Specification;
Catalogue number 5041 Calcium Flow PlusCombination ISE
Parameter
Specification
Overall length
Body Diameter
Cap Diameter
Connector
Cable length
Resistance at 25 Deg C
Concentration Range
Slope
Potential Drift
Operating pH range
Temperature range
Endpoint time
Interferences. Ions with coefficients above 0.001.
155 mm
12 mm
16mm
BNC
180 mm
< 2.5 Meg Ohm
0.02 to 40100 ppm
24 to 29 mV per decade
2 mV per day
3.5 to 11
5 to 50 Deg C
Typically 10 to 30 seconds
Strontium, Barium, Copper
Price: £395.00
Files to download Applications advice. Choose the right ISE 106.9 KB Flow Plus ISE Brochure 320.7 KB