» Ammonium Combination ISE 3051
The directION Ammonium combination ion selective electrode has a solid state PVC polymer membrane with an integral driTEK reference electrode. The electrode is designed for the detection and analysis of Ammonium ions in aqueous solutions and is suitable for use in the field in the laboratory and in on line analyzers.
Specification
Catalogue number 3051 Ammonium directION Combination ISE
Parameter
Overall length
Body Diameter
Cap Diameter
Connector
Cable length
Resistance at 25 Deg C
Concentration Range
Slope
Potential Drift
Operating pH range
Temperature range
Endpoint time
Interferences: Ions with coefficients above 0.001.
155 mm
12 mm
16mm
BNC
1000 mm
< 2.5 Meg Ohm
0.9 to 9,000 ppm
54 to 59 mV per decade
2 mV per day
0 to 8.5
5 to 50 Deg C
Typically 10 to 30 seconds
Calcium, Potassium, Sodium
Files to download Ammonium Combination Instructions 194.4 KB Applications advice. Choose the right ISE 106.9 KB