» Lead Combination ISE 3231
The directION Lead combination ion selective electrode has a solid state crysytalline membrane with an integral driTEK reference electrode. The electrode is designed for the detection and analysis of Lead ions in aqueous solutions and is suitable for use in the field in the laboratory and in on line analyzers.
Specification
Catalogue number 3231 Lead directION Combination ISE
Parameter
Overall length
Body Diameter
Cap Diameter
Connector
Cable length
Resistance at 25 Deg C
Concentration Range
Slope
Potential Drift
Operating pH range
Temperature range
Endpoint time
Interferences: Ions with coefficients above 0.001.
155 mm
12 mm
16mm
BNC
1000 mm
< 2.5 Meg Ohm
0.2 to 20,800 ppm
24 to 29 mV per decade
2 mV per day
3 to 7
5 to 50 Deg C
Typically 20 to 50 seconds
Mercury, Silver, Copper, Cadmium
Files to download Applications advice. Choose the right ISE 106.9 KB Lead Combination ISE Instructions 39 KB