Automatic Temperature Compensation and ISE's?


Q Can I do Automatic Temperature compensation with Ion Selective Electrodes like I do with my pH meter?

A It is good practice to record the temperature when doing ISE work but you cannot perform ATC like you do with pH. Best practice is to make sure your standards and samples are at the same temperature. 

It is possible to make a manual correction estimate after you have the concentration on your meter if you know the difference between your sample and standard temperature. This is by use of the Nernst equation. Please ask for details.